21 - 6
21.2.7.4
Viewing Event Details
To view waveforms and parameter values at the event time, follow this procedure:
1.
Enter the
Events
page.
2.
Select
Overview
.
To display beat labels on the first ECG waveform, switch on
Beat Annotation
. The white beat labels indicate
heart beats classification and may explain suspected, missed, or false arrhythmia calls. Heart beats are classified
as follows:
■
N = Normal
■
V = Ventricular ectopic
■
S = Supraventricular premature
■
P = Paced
■
L = Learning
■
? = Insufficient information to classify beats
■
I = Inoperative (for example, Lead Off )
■
M = Missed beat
If you switch on
Beat Annotation
on the
Events
page, beat labels will also be displayed on the
Full Disclosure
page, and vise versa. Beat labels can be printed out.
21.2.7.5
Printing Event Reports
You can print event reports either via a printer or via a recorder.
To do so, follow this procedure:
1.
Enter the events review page.
2.
Select
to enter the
Print Setup
menu.
3.
Select the desired options.
◆
Print Event List
: print the entire event list.
◆
Print List of Selected Events
: print the list of selected events.
◆
Print Detail of Selected Events
: print the details of selected events.
◆
Print Displayed Event Detail
: print the waveforms and parameters of the currently displayed event.
4.
To print a report via a recorder, select
.
21.2.8
Full Disclosure Review Page
You can review up to 48-hour waveform data on the full disclosure review page. You can view both the
compressed waveforms, full waveforms and numeric values.
21.2.8.1
Entering the Full Disclosure Review Page
Choose one of the following methods to enter the full disclosure review page:
■
Select the
Review
quick key
→
select the
Full Disclosure
tab.
■
Select the
Main Menu
quick key
→
from the
Review
column select
Full Disclosure
.
21.2.8.2
Selecting Waveforms
Before reviewing compressed waveforms, you must select waveforms you want to store and display. To store and
display the desired waveforms, follow this procedure:
1.
Enter the full disclosure review page.
2.
Select
Setup
to enter the
Select Waveform
page.
3.
Select the
Storage
tab and set the desired waveforms to be stored in the monitor. Select the
Display
(Maximum: 3)
tab and set the desired waveforms to be displayed on the
Full Disclosure
page.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...