12 - 4
If you cannot get optimal Resp waveform or you suspect the Resp value after choosing the Resp lead, you may
need to optimize the electrode placement.
12.5.4
Setting the Resp Waveform Size
To set the Resp waveform size, follow this procedure:
1.
Select the Resp numeric area or waveform area to enter the
Resp
menu.
2.
Select the
Setup
tab.
3.
Set
Gain
.
12.5.5
Setting the Resp Waveform Speed
To set the Resp waveform speed, follow this procedure:
1.
Select the Resp numeric area or waveform area to enter the
Resp
menu.
2.
Select the
Setup
tab.
3.
Set
Speed
.
12.5.6
Setting the Auto Detection Switch
To set the auto detection switch, follow this procedure:
1.
Select the Resp numeric area or waveform area to enter the
Resp
menu.
2.
Select the
Setup
tab.
3.
Switch on or off
Auto Threshold Detection
.
◆
If
Auto Threshold Detection
is switched on, the monitor automatically adjusts the Resp waveform
detection level, or threshold.
◆
If
Auto Threshold Detection
is switched off, you have to manually adjusts the Resp waveform
threshold. For more information, see
12.5.7 Adjusting the Resp Waveform Detection Threshold
In the auto detection mode, if you are monitoring Resp and ECG is switched off, the monitor cannot compare the
ECG and Resp rates to detect cardiac overlay. The respiration detection level is automatically set higher to
prevent the detection of cardiac overlay as respiration.
In the manual detection mode, cardiac overlay can in certain situations trigger the respiration counter. This may
lead to a false indication of a high respiration or an undetected apnea condition. If you suspect that cardiac
overlay is being registered as breathing activity, raise the detection level above the zone of cardiac overlay. If the
Resp wave is so small that raising the detection level is not possible, you may need to optimize the electrode
placement.
12.5.7
Adjusting the Resp Waveform Detection Threshold
Use the manual detection mode in the following situations:
■
The respiration rate and the heart rate are close.
■
Patients have intermittent mandatory ventilation.
■
Respiration is weak. Try repositioning the electrodes to improve the signal.
To set the Resp waveform threshold to the desired level, follow this procedure:
1.
Select the Resp numeric area or waveform area to enter the
Resp
menu.
2.
Select the
Threshold
tab.
3.
Select the up and down arrows below
Upper Line
and
Lower Line
to define the Resp waveform threshold.
Once set, the detection level will not adapt automatically to different respiration depths. It is important to
remember that if the depth of breathing changes, you may need to change the detection level.
12.6
Resp Troubleshooting
For more information, see
.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...