13 - 8
4.
Set the alarm properties as desired.
13.7.2
Changing the QRS Volume
If the
Alarm Source
is set to
PR
, the QRS tone is derived from PR measurements. To set the QRS volume, follow
this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
PR
tab.
3.
Select the
Setup
tab.
4.
Set
QRS Volume
.
If the SpO
2
value is effective, the monitor also adjusts the QRS tone (pitch tone) according to the SpO
2
value. For
information, see
.
13.7.3
Setting the PR Source
Current pulse source is displayed in the PR numeric area. The PR from current pulse source has the following
characteristics:
■
PR is monitored as system pulse and generates alarms when you select PR as the active alarm source.
■
PR is stored in the monitor’s database and reviewed in the graphic/tabular trends; in trend graphs, as the PR
curve is in the same color with that of the PR source, it is unlikely to distinguish the PR source.
■
PR is sent via the network to the CMS, if available.
To set which pulse rate as PR source, follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
PR Setup
tab.
3.
Set
PR Source
.
The
PR Source
menu displays the currently available PR sources from top to bottom by priority. When you select
Auto
, the system will automatically select the first option as the PR source. If the current PR source is unavailable,
the system will automatically switch
PR Source
to
Auto
. When you select
IBP
, the system will automatically
select the first pressure label as the PR source.
13.7.4
Showing/Hiding PR
You can set whether to display the PR value in the SpO
2
parameter area. To do so, follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
PR
tab.
3.
Select the
Setup
tab.
4.
Switch on or off
Display PR.
13.8
Displaying SpO
2
Statistics
You can show SpO
2
statistics for a defined period of time. To do so, follow this procedure:
1.
Access the
Tile Layout
in either of the following ways:
◆
Select the
Screen Setup
quick key
→
select the
Tile Layout
tab.
◆
Select
the
Main Menu
quick key
→
from the
Display
column select
Tile Layout
.
2.
Select the parameter numeric area where you want to display SpO
2
statistics, and then from the popup list
select
SpO2
→
SpO2 Statistics
.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...