13 - 9
The following figure shows the SpO
2
statistics area.
13.8.1
Selecting the Range of each SpO
2
Section and the Target Section
To define the SpO
2
range of each section, follow this procedure:
1.
Select the SpO
2
statistics area.
2.
From the
To
column select the SpO2 value at which corresponding section ends.
3.
From the
Target
column select the target section. The target section is highlighted in green in the SpO
2
statistics area.
13.8.2
Selecting the SpO
2
Statistics Length
The Duration of SpO
2
statistics is configurable. From the SpO
2
statistics area, select the duration to redefine the
length of SpO
2
statistics.
13.9
SpO
2
Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or
accessories, check the table below before requesting for services. If the problem persists, contact your service
personnel.
NOTE
•
For the physiological and technical alarm messages, see
.
(1)
Duration of SpO
2
statistics
(2)
Results of SpO
2
statistics
(3)
Sections for statistics: The section in green indicates the target range.
(1)
(2)
(3)
Problem
Solution
Do not see SpO
2
numeric area or waveform area
on the main screen
1. Check that the SpO
2
is set to display in the
Screen Setup
menu.
For more information, see
2. Check that if the SpO
2
parameter switch is enabled. If not, enable
the SpO
2
measurement
.
For more information, see
.
3. Check that the cable connections of SpO
2
sensor and the extension
cable are tight. Replace the SpO
2
sensor or the extension cable if
needed.
Dashes “- -” display in place of numerics.
1. Check that the cable connections of SpO
2
sensor and the
extension cable are tight. Replace the SpO
2
sensor or the extension
cable if needed.
2. Reconnect the SpO
2
sensor if the alarm
SpO2 Sensor Off
appears.
3. Check the PI value. If the PI value is too low, adjust the SpO
2
sensor,
or apply the sensor to the site with better perfusion.
4. Move the sensor to the place with weaker light, or cover the sensor
with shade cloth if the alarm
SpO2 Sensor Off
appears.
Low amplitude SpO
2
signal
1. The SpO
2
sensor and NIBP cuff are placed on the same limb.
Change a monitoring site if necessary.
2. Check the PI value. If the PI value is too low. Adjust the SpO
2
sensor,
or apply the sensor to the site with better perfusion.
3. Check the sensor and its application site.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...