16 - 4
If you have to transfer the patient who is taking ICP measurement, check that the target monitor supports the
Codman ICP transducer. For more information, see
16.4.1 Zeroing the Codman ICP transducer
. If the target
monitor does not support the Codman ICP transducer, do not use it for ICP monitoring.
Follow this procedure to transfer the patient:
1.
Disconnect the ICP adapter cable from the measurement module, or remove the module from the monitor.
2.
Connect the ICP adapter cable, measurement module, and the target monitor, or insert the measurement
module into the target monitor.
3.
Check that the zero reference value displayed on the monitor is consistent with that recorded on the ICP
transducer.
◆
Consistent: select
Accept
.
◆
Inconsistent: input the zero reference value recorded on the ICP transducer, and select
Accept
.
CAUTION
•
If monitors of different brands are used to zero the Codman ICP transducer, the zero reference
values can be different. Use a Mindray monitor to Zero the Codman ICP transducer if you will take
ICP measurement using a Mindray monitor. Otherwise the ICP measurement can be inaccurate.
16.5
IBP Display
The IBP measurement is displayed on the monitor as a waveform and numeric pressures. For arterial pressure,
the IBP numeric area displays systolic pressure, diastolic pressure and mean pressure. For venous pressure, the
IBP numeric area displays only the mean pressure. The figure below shows the waveform and numerics for the
Art pressure.
16.6
Changing IBP Settings
16.6.1
Changing the IBP Alarm Settings
To change the IBP alarm settings, follow this procedure:
1.
Select the IBP numeric area or waveform area to enter the corresponding pressure menu.
2.
Select the
Alarm
tab.
3.
Enter the password if required.
4.
Set the alarm properties.
(1) Pressure label
(2) Waveform
(3) Pressure Unit
(4) Systolic pressure
(5) Diastolic pressure
(6) Mean pressure
(7) PPV measurement
(7) (6)
(1) (2) (3) (4) (5)
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...