B - 2
NOTE
•
If the essential performance is lost or degraded, it may be necessary to take mitigation measures,
such as re-orienting or relocating the ME EQUIPMENT or ME SYSTEM or shielding the location or
stopping using the monitor and contact the service personnel.
•
The device needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided below.
•
Other devices may interfere with this device even though they meet the requirements of CISPR.
•
When the inputted signal is below the minimum amplitude provided in technical specifications,
erroneous measurements could result.
•
The EMISSIONS characteristics of this device make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required) this device might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or re-orienting the
device.
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Voltage dips and
voltage interruptions
IEC 61000-4-11
0% U
T
for 0.5 cycle: at
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% U
T
for 1 cycle and
70% U
T
for 25/30
cycles: at 0°
0% U
T
for 250/300
cycle
0% U
T
for 0.5 cycle: at
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% U
T
for 1 cycle and
70% U
T
for 25/30
cycles: at 0°
0% U
T
for 250/300
cycle
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of our product requires continued
operation during power mains interruptions,
it is recommended that our product be
powered from an uninterruptible power
supply or a battery.
RATED power
frequency magnetic
fields
IEC 61000-4-8
30 A/m
50 Hz/60 Hz
30 A/m
50 Hz/60 Hz
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...