D - 9
D.2.11
BIS Technical Alarm Messages
Note: XX represents BIS label, for example G, C, LE, LT, RL-RA, L-R, F-R, 1, 2, and so on.
D.2.12
EWS Technical Alarms
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
BIS Sensor Off
Low
A
Check and reconnect the BIS sensor. If the alarm
persists, replace the sensor.
BIS Electrode XX Off
Low
A
Check the electrode connection, and re-attach
the electrodes if necessary.
BIS Electrode XX Poor
Contact
Low
A
Enter the
Sensor Check
menu, and check the
connections of the sensor and electrodes.
BISx Error
High
C
Replug the module. If the alarm persists, contact
your service personnel.
BIS No Sensor
Low
A
BIS sensor is not properly connected, BIS cable
fails, BISx or BISx4 fails.
1. Check the BIS sensor connection.
2. Replug the BIS module.
3. Replace the BIS cable.
4. Replace BISx or BISx4.
BIS Sensor Too Many Uses
Low
A
Replace the sensor.
BIS Sensor Old
Low
A
Replace the sensor.
BIS XX Signal Quality Too
Low
Low
A
SQI < 15
1. Check the patient’s condition.
2. Check the sensor position, and its contact with
the patient’s skin.
3. Check that BISx or BISx4 is away from the
electrically radiating equipment.
BIS XX Low Signal Quality
Low
A
SQI < 15
1. Check the patient’s condition.
2. Check the sensor position, and its contact with
the patient’s skin.
3. Check that BISx or BISx4 is away from the
electrically radiating equipment.
BIS Wrong Sensor Type
Low
A
Check or replace the sensor.
BIS Sensor Error
Low
C
Replace the sensor.
BISx Disconnected
Low
A
BISx or BISx4 is not connected properly or fails, or
BIS patient cable fails.
1. Check that BISx or BISx4 is connected properly.
2. Replug the BIS Module.
3. Replace the BIS patient cable.
4. Replace BISx or BISx4.
Reconnect BISx
Low
A
Replug the BIS Module.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
EWS param XX is timeout
Low
A
The manually input parameter is timeout. Input a
parameter numeric again.
EWS score needs to be
confirmed
Low
A
Confirm to save or give up current score.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...