27 - 6
27.4.5
The Nurse Call Tab
NOTE
•
The call help function works only when the monitor is connected to the network.
•
The call help sound may disturb patients in the same department.
27.4.6
The Other Tab
Menu Item
Default Setting
Description
Signal Type
Continuous
•
Pulse
: the nurse call signal is a pulse signal
and each pulse lasts one second. When
multiple alarms simultaneously occur, only
one pulse signal is outputted. If an alarm
occurs but the previous one is not cleared, a
new pulse signal will also be outputted.
•
Continuous
: the nurse call signal lasts until
the alarm ends. That is to say the duration of a
nurse call signal is equal to that of the alarm
condition.
Contact Type
Normally Open
Selects the work mode of the nurse call relay
Alarm Priority
High Only
Selects the priority of alarms sent to the nurse call
system
Alarm Type
Physiological Only
Selects the type of alarms sent to the nurse call
system.
Receive Call Help
On
Receives the calling signal if a monitor in the
same department calls for help.
Section
Menu Item
Default Setting
Description
Alarm Priority
ECG Lead Off
Low
Selects the priority of the ECG lead off alarm.
SpO2 Sensor Off
Low
Selects the alarm level for SpO
2
sensor off alarm.
IBP No Sensor
Med
Selects the alarm level for IBP No Sensor alarm.
CMS/eGW
Disconnected
Low
Selects the priority of the CMS and eGateway
disconnection alarm.
Alarm Delay
Alarm Delay
6 sec
•
1 sec ~15 sec
: for continuously measured
parameters, an alarm is not presented if the
alarm condition is resolved within the
designated delay time.
•
Off
: an alarm is always presented.
The setting of
Alarm Delay
is not applied to the
apnea alarms and the ST alarms.
ST Alarm Delay
30 sec
The monitor does not present the ST alarm if the
alarm condition is resolved within the delay time.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...