A - 19
A.13.9
CO
2
Specifications
Sidestream CO
2
Module
Measurement mode
Sidestream, microstream, mainstream
Technique
Infrared absorption
Apnea time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Alarm limit
Range
Step
EtCO
2
High
(low limit + 2) to 99 mmHg
1 mmHg
EtCO
2
Low
1 to (high limit - 2)mmHg
FiCO
2
High
1 to 99 mmHg
EtO
2
High
(low limit + 2%) to 100%
1%
EtO
2
Low
0% to (high limit - 2)%
FiO
2
High
(low limit + 2%) to 100%
FiO
2
Low
18% to (high limit - 2)%
Standard
Meet the standard of ISO 80601-2-55
CO
2
Measurement range
0 to 150 mmHg
CO
2
absolute accuracy*
Full accuracy mode:
0 to 40 mmHg: ± 2 mmHg
41 to 76 mmHg: ±5% of reading
77 to 150 mmHg: ±10% of reading
ISO accuracy mode: add ±2mmHg to the full accuracy mode
Inaccuracy specifications are affected by the breath rate and I:E. The EtCO
2
accuracy is within specification for breath rate ≤
60 rpm and I/E ratio ≤ 1:1, or breath rate ≤ 30 rpm and I/E ratio ≤ 2:1.
CO
2
resolution
1 mmHg
Recovery time
<15 s (after defibrillation)
O
2
measurement range
0 to 100%
O
2
absolute accuracy
0≤O
2
concentration≤25%: ±1%
25<O
2
concentration≤80%: ±2%
80<O
2
concentration≤100%: ±3%
O
2
resolution
1%
Accuracy drift
Meet the requirement for measurement accuracy within 6 hours
Sample flowrate
For sidestream CO
2
module without O
2
monitoring function:
Connected a DRYLINE II watertrap for adult and pediatric patient: 120 ml/
min
Connected a DRYLINE II watertrap for neonatal patient: 90 ml/min or 70 ml/
min
For sidestream CO
2
module with O
2
monitoring function:
Connected a DRYLINE II watertrap for adult and pediatric patient: 120 ml/
min
Connected a DRYLINE II watertrap for neonatal patient: 90 ml/min
Sample flowrate tolerance
±15% or ±15 ml/min, whichever is greater.
Start-up time
Maximum: 90 s
Typically: 20 s
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...