20 - 6
◆
If you set
Display
to EEG waveforms (
EEG LT
or
EEG LE
), set
Scale
and
Speed
for EEG waveforms.
◆
If you set
Display
to BIS parameter trends, set
Trend Length
.
20.7.4
Switching Off the Filter
The filter can filter EEG interference. It is switched on by default.
To disable the filter, follow this procedure:
1.
Select the BIS numeric area or waveform area to enter the
BIS
menu.
2.
In the
BIS
menu, select the
Setup
tab.
3.
Switch off
Filter
.
20.7.5
Setting the Displayed BIS Parameters
Besides the BIS value, you can also display up to four secondary parameters in the BIS numeric area. To select the
displayed parameters, follow this procedure:
1.
Select the BIS numeric area or waveform area to enter the
BIS
menu.
2.
In the
BIS
menu, select the
Select Parameter
tab.
3.
From the
BIS Tile
area, select a secondary parameter block, and then select a secondary parameter from
the
Parameters
area.
20.8
Sensor Check
20.8.1
Automatic Sensor Check
Once the BIS sensor is connected, an automatic sensor check starts to check the sensor type, status, and
impedance of all the electrodes, including the signal electrodes, the reference electrode and the ground
electrode. During the sensor check, the message “Sensor Check In Progress” displays in the information area. If
this message continuously displays, enter the sensor check menu and check if the impedance of each electrode
is acceptable.
After the initial sensor check, the monitor performs automatic check during BIS monitoring. Automatic sensor
check includes the following items:
■
The combined impedance of the signal electrodes plus the reference electrode. This is done automatically
and continuously and does not affect the EEG waveform. As long as the impedances are within the valid
range, there is no prompt message of this check or its results.
■
The impedance of the ground electrode. This is done every ten minutes and takes approximately four
seconds. It causes artifact in the EEG waveform, and the message
BIS: Checking Ground
is displayed on
the monitor during the check. If the ground electrode does not pass this check, another check is initiated.
This continues until the ground electrode passes the check.
The monitor continually checks impedance levels during a procedure by generating a 128 Hz test signal.
Occasionally this signal may interfere with other equipment. If this becomes a problem, you may need to disable
the auto checking. Impedance levels will still be tested at startup, but once they pass, they will not be tested
again until a new case is begun.
20.8.2
Disabling Automatic Sensor check
To disable automatic sensor check, follow this procedure:
1.
Select the BIS numeric area or waveform area to enter the
BIS
menu.
2.
Select the
Setup
tab.
3.
Switch off
Auto Check
.
CAUTION
•
Automatic sensor check may need to be disabled if the 1 nA 128 Hz impedance check signal
interferes with other equipment.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...