9 - 9
■
Priorities of paused alarms
■
Alarm reset setting
■
Reminder tone settings
For more information, see
.
9.7.2.1
Switching Off All Alarms
If
Pause Time
is set to
Permanent
(see
), pressing the
Alarm Pause
quick key
permanently switches off all alarms. The alarm off status has the following features:
■
Physiological alarms are switched off. The alarm lamp does not flash and alarm sound is not issued.
■
Alarm sound of technical alarms is switched off, but alarm lamp flashes and alarm messages are presented.
■
The message
Alarm Off
with red background is displayed in the physiological alarm information area.
■
The alarm off symbol is displayed in the system status information area.
To exit the alarm off status, press the
Alarm Pause
quick key again.
WARNING
•
Pausing or switching off alarms may result in a hazard to the patient.
9.7.3
Pausing Alarm Sound
If the pause function is defined as
Audio Pause
, pressing the
Audio Pause
key pauses alarm tone. When alarm
tones are paused, the following rules are followed:
■
The sound of all physiological alarms and technical alarms are switched off.
■
The remaining audio pause time is displayed in the physiological alarm information area.
■
The audio pause symbol is displayed in the system information area.
When the audio pause time expires, the audio paused status is automatically deactivated. You can also cancel
the audio paused status by pressing the
Audio Pause
quick key.
9.7.3.1
Setting the Alarm Tone Pause Time
The alarm tone pause time can be set to
1 min
,
2 min
,
3 min
, or
Permanent
. The default audio pause time is two
minutes.
This function is password protected. For more information, see
.
9.7.3.2
Prolonging the Alarm Tone Pause Time
You can temporarily prolong the alarm tone pause time after the monitor enters the alarm tone paused status.
This function is password protected. For more information, see
.
NOTE
•
Prolonging alarm pause time does not affect the setting of alarm tone pause time.
9.7.3.3
Setting the Priority of Audio Paused Alarms
You can select alarm sound of what priority can be paused.This function is password protected. For more
information, see
9.7.3.4
Switching Off Alarm Sound
If
Pause Time
is set to
Permanent
(see
), pressing the
Audio Pause
quick key
permanently switches off all alarm sound. The audio off status has the following features:
■
Alarm sound of both physiological alarms and technical alarms is switched off.
■
The audio off symbol is displayed in the system information area.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...