23 - 3
23.4
Hemodynamic Calculations
The monitor provides the hemodynamic calculation function. The monitor can save the results of up to 10
calculations, which are displayed in groups.
23.4.1
Performing Hemodynamic Calculations
To perform hemodynamic calculation, follow this procedure:
1.
Access hemodynamic calculation by either of the following ways:
◆
Select the
Calculations
quick key
→
Hemodynamics
tab.
◆
Select the
Main Menu
quick key
→
from the
Calculations
column
select
Hemodynamics
.
2.
Enter the known values. For a patient who is being monitored, the currently measured values are automat-
ically taken.
3.
Select
Calculate
.
The calculated value greater than the normal upper limit is indicated by an up arrow “
↑
”. The calculated value
lower than the normal lower limit is indicated by a down arrow “
↓
”.
You can select
Range
to show the normal range of each parameter.
23.4.2
Input Parameters for Hemodynamic Calculations
NOTE
•
If you enable Use PA-D as PAWP, PA-D value will be used to replace PAWP value for hemodynamic
calculation. For more information, refer to
16.6.8 Setting the Use PA-D as PAWP Switch
.
23.4.3
Calculated Parameters and Formulas for Hemodynamic Calculations
Input Parameter
Label
Unit
cardiac output
C.O.
L/min
heart rate
HR
bpm
pulmonary artery wedge pressure
PAWP
mmHg
artery mean pressure
PMAP
mmHg
pulmonary artery mean pressure
PA Mean
mmHg
central venous pressure
CVP
mmHg
end-diastolic volume
EDV
ml
height
Height
cm
weight
Weight
kg
Calculated Parameters
Label
Unit
Formula
cardiac index
C.I.
L/min/m
2
C.I. (L/min/m
2
) = C.O. (L/min)/BSA (m
2
)
body surface area
BSA
m
2
BSA (m
2
) = Wt
0.425
(kg) × Ht
0.725
(cm) × 0.007184
stroke volume
SV
ml
SV (ml) = 1000× C.O. (L/min)/HR (bpm)
stroke index
SVI
ml/m
2
SVI (ml/m
2
) = SV (ml)/BSA (m
2
)
systemic vascular
resistance
SVR
DS/cm
5
SVR (DS/cm
5
) = 79.96 × [PAMAP (mmHg) - CVP
(mmHg)]/C.O. (L/min)
systemic vascular
resistance index
SVRI
DS•m
2
/cm
5
SVRI (DS•m
2
/cm
5
) = SVR (DS/cm
5
) × BSA (m
2
)
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...