A - 24
Response time
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal
sampling line:
120 ml/min:
CO
2
:
N
2
O:
O
2
:
Hal, Iso, Sev, Des, Enf:
≤4 s
≤4.2 s
≤4 s
≤4.4 s
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling
line:
200 ml/min:
CO
2
:
N
2
O:
Hal, Iso, Sev, Des, Enf:
O
2
:
≤4.2s
≤4.3s
≤4.5s
≤4s
Anesthetic agent limit
Primary anesthetic agent
In full accuracy mode: 0.15%,
Secondary anesthetic agent:
In full accuracy mode: 5% of primary agent if primary agent is greater than
10%, 0.3% if primary agent is less than or equal to 10%.
Data sample rate
25 Hz
Inaccuracy specifications are affected by the breath rate and I:E change. The end-tidal gas reading is within specification for
breath rate below 15BPM and I:E ratio smaller than 1:1 relative to the gas readings without breath; Add ±6%REL to
inaccuracy for HAL and O
2
for breath rate larger than 15 BPM; Add ±6%REL to inaccuracy for all gases for breath rate larger
than 30 BPM (inaccuracy for HAL and O
2
are unspecified in this case); inaccuracy is unspecified for breath rate larger than 60
BPM.
Effect of interference gases on AG measurements
Gas
Concentration
(%)
Quantitative effect(%ABS)3)
CO
2
N
2
O
Agent
1)
O
2
CO
2
/
/
0.1
0
0.2
N
2
O
/
0.1
/
0.1
0.2
Agent
1) 2)
/
0.1
0.1
0.1
1
Xenon
<100%
0.1
0
0
0.5
Helium
<50%
0.1
0
0
0.5
Ethanol
<0.1%
0
0
0
0.5
Acetone
<1%
0.1
0.1
0
0.5
Methane
<1%
0.1
0.1
0
0.5
Saturated Isopropanol vapour
/
0.1
0
0
0.5
Metered dose inhaler propellants,
/
Unspecifie
d
Unspecifie
d
Unspecifie
d
Unspecifie
d
O
2
/
0.2
0.2
1.0
/
1) Agent represents one of Des, Iso, Enf, Sev, and Hal.
2) Multiple agent interference on CO
2
, N
2
O and O
2
is typically the same as single agent interference.
3) For CO
2
, N
2
O and Agents, maximum interference from each gas at concentrations within specified accuracy ranges for
each gas. The total interference of all gases is never larger than 5%REL.
Alarm limit
Range
Step
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...