19 - 2
19.2
AG Safety Information
WARNING
•
To avoid explosion hazard, do not use flammable anesthetic agent such as ether and cyclopropane
for this equipment.
•
The presence of other substances in the patient’s breathing circuit, such as ethanol, acetone,
methanol, isopropanol, freon, asthma medication carrier gases, and other infrared absorbing gases,
can influence the anesthesia agent identification and lead to incorrect measurements and
identification.
•
Using high-frequency electrosurgical units may increase the risk of skin burn. In this case, do not use
antistatic or conductive respiratory tubing.
•
Route all tubing away from the patient’s throat to avoid strangulation.
CAUTION
•
Perform the measurement in a well-ventilated environment.
•
EtCO
2
values measured from the AG module may differ from that of from the blood gas analysis.
NOTE
•
The AG module automatic suppress physiological alarms until breathing waves have been detected.
Make sure that a patient is properly connected when monitoring with the AG module.
19.3
AG Measurement Limitations
The following factors may influence the measurement accuracy:
■
Leaks or internal venting of sampled gas
■
Mechanical shock
■
Cyclic pressure up to 10 kPa (100 cmH
2
O)
■
Other sources of interference, if any
19.4
AG Display
The AG module can send waves and numerics for all measured anesthetic gases for display on the monitor,
including:
■
CO
2
, O
2
, N
2
O and AA waves
■
awRR: airway respiratory rate
■
MAC: minimum alveolar concentration
■
End tidal (Et) and fraction of inspired (Fi) numerics for CO
2
, O
2
, N
2
O and AA
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...