C - 12
C.1.7
C.O. Default Settings
C.1.8
CO
2
Default Settings
C.1.8.1
General Settings
Display Format
Sys/Dia (Mean)
Display Alarm Limit
Off
Use PA-D as PAWP
Off
Item
Default Setting
Item
Default Setting
TB
Alarm switch
On
High limit
39.0 °C
Low limit
36.0 °C
Priority
Med
Alarm Outputs
Off
Comp Const
0.542
Auto Start
On
Auto TI
On
Item
Default Setting
EtCO2 Alarm
switch
On
High limit
Adult, Pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit
Adult, Pediatric: 25mmHg
Neonate: 30mmHg
Priority
Med
Alarm Outputs
Off
FiCO2 Alarm
switch
On
High limit
4 mmHg
Priority
Med
Alarm Outputs
Off
Apnea Delay
Adult, Pediatric: 20 s
Neonate: 15 s
RR Source
Auto
Speed
6.25 mm/s
Scale
50 mmHg
Waveform Type
Draw
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...