15 - 4
d
Middle of the cuff should be at the level of the right atrium of the heart. If it is not, you must use the
measurement correction formula to correct the measurement. For more information, see
15.9.10 Correcting the NIBP Measurements
4.
Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass
unrestricted through the tubing.
CAUTION
•
A wrong cuff size and a folded or twisted bladder can cause inaccurate measurements.
•
Do not touch or apply external pressure against the cuff and air tubing during NIBP measurement.
This may cause inaccurate blood pressure values.
•
Use care when placing the cuff on an extremity used for monitoring other patient parameters.
15.7
Starting and Stopping NIBP Measurements
Start and stop NIBP measurement by selecting the NIBP quick keys or from the NIBP menu.
15.8
Viewing NIBP Analysis
NIBP analysis provides you a dynamic analysis of NIBP changes and distribution over the time scale. It allows you
to know the patient’s condition of the latest 24 hours before you entering the NIBP Analysis window.
To view NIBP analysis, follow this procedure:
1.
Select the NIBP numeric area to enter the
NIBP
menu.
2.
Select the
Analysis >>
tab.
You can also select anywhere in the
Analysis
window to enter the
Review
window. For more information, see
.
Task
By Quick Key
From NIBP menu
Start a manual measurement
NIBP Start/Stop
quick key
Start NIBP
button
Start auto NIBP series
NIBP Start/Stop
quick key
Make sure to set
Interval
before
starting auto NIBP.
Setup
tab
→
set
Interval
→
Start NIBP
button
NIBP Measure
quick key
→
select
Interval
Start
NIBP sequence measurement
NIBP Measure
quick key
→
Sequence
Sequence
tab
→
set NIBP sequence
→
Start NIBP
button
Start STAT measurement
NIBP STAT
quick key
STAT
button
NIBP Measure
quick key
→
STAT
Stop the current NIBP
measurements
NIBP Start/Stop
quick key
Stop NIBP
button
End auto NIBP series or NIBP
Sequence
NIBP Stop All
quick key
NIBP Stop All
button
Stop STAT measurement and end
series
NIBP Start/Stop
quick key
Stop NIBP
or
NIBP Stop All
button
NIBP Stop All
quick key
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...