21 - 7
NOTE
•
The more waveforms selected in the Storage column, the shorter the waveform storage time. The
waveforms may not be stored for 48 hours. Please exert caution when selecting waveforms.
In case of alarms, the background of compressed waveform block at the alarm time is marked with a special
color:
■
Red: high alarm priority
■
Yellow: medium alarm priority
■
Cyan: low alarm priority
21.2.8.3
Setting Scale and Duration
To set the length and size of displayed compressed waveforms, follow this procedure:
1.
Enter the full disclosure review page.
2.
Select
...,
and then select
Scale
to set ECG waveform gain.
3.
Select
Duration
to set the length of displayed waveforms.
4.
Select parameter label to set scale for each parameter.
21.2.8.4
Viewing Details of Compressed Waveforms
To view the full waveforms and numeric values, follow this procedure:
1.
Enter the full disclosure review page.
2.
Select
Detail
.
You can perform the following operations on the this page:
■
Switch on
Beat Annotation
. For more information, see
21.2.7.4 Viewing Event Details
■
Select
...
and set
Speed
and
ECG Gain
, or
Save As Event
.
■
Select
Overview
to switch to the compressed waveform page.
21.2.8.5
Printing the Full Disclosure Waveform Report
To print a compressed waveform
report, follow this procedure:
1.
Enter the full disclosure review page.
2.
Select
and set the time range for printing.
21.2.9
OxyCRG Review Page
You can review up to 48 hours of 4-minute trend curves on the OxyCRG review page. The OxyCRG review
functionality is applicable for neonatal monitoring only.
21.2.9.1
Entering the OxyCRG Review Page
Choose one of the following methods to enter the OxyCRG review page:
■
From the OxyCRG screen, select the ABD events list area.
■
Select the
Review
quick key
→
select the
OxyCRG
tab.
■
Select the
Main Menu
quick key
→
from the
Review
column select
OxyCRG
.
NOTE
•
OxyCRG Review Page is available only when Patient Category is set to Neo.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...