D - 5
D.2.3
Resp Technical Alarm Messages
D.2.4
SpO
2
Technical Alarm Messages
D.2.5
Temp Technical Alarm Messages
Note: XX represents a temperature site, for example skin, core, axil, T1, and so on.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
Resp Interference
Prompt
/
The respiration circuit is disturbed. Check for any
possible sources of signal noise.
Electrode Poor Contact
Prompt
/
Check the electrode application. Reposition or
replace the electrodes if necessary.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
SpO2 Sensor Off
Adjustable
B
The SpO
2
sensor has become detached from the
patient or the module. Check the sensor
connection. If the alarm persists, replace the
sensor.
SpO2 No Sensor
Low
A
The SpO
2
extension cable is detached from the
SpO
2
module, or the SpO
2
sensor is detached
from the SpO
2
extension cable. Check the SpO
2
cable and the sensor connection. If the alarm
persists, replace the sensor.
SpO2 Excess Light
Low
C
Ambient light is too strong. Move the sensor to a
place with lower level of ambient light or cover
the sensor to minimize the ambient light.
SpO2 No Pulse
Low
C
The SpO
2
sensor failed to obtain pulse signal.
Check the patient’s condition and replace the
sensor application site. If the alarm persists,
replace the sensor.
SpO2 Sensor
Incompatible
Low
C
Incompatible or an unspecified SpO
2
sensor is
used. Use specified sensors.
SpO2 Low Signal Quality
Low
C
1. Check the sensor and sensor position.
2. Make sure the patient is not shivering or
moving.
3. The patient’s pulse may be too low to be
measured.
SpO2 Interference
Low
C
The SpO
2
signal has been interfered. Check for
any possible sources of signal noise and check
the patient for excessive motion.
SpO2 Sensor Error
Low
C
Replace the sensor and measure again.
SpO2 Searching Pulse
Prompt
/
SpO
2
is searching for pulse.
SpO2 Low Perfusion
Prompt
/
The SpO
2
sensor is not properly placed or the
patient’s perfusion index is too low.
1. Check the sensor and sensor position.
2. Reposition the sensor if necessary.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
T XX Sensor Off
Low
A
Check the sensor connection and reconnect the
sensor.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...