10 - 10
◆
Surgery
: use when the signal is distorted by high frequency or low frequency interference. High
frequency interference usually results in large amplitude spikes making the ECG signal look irregular.
Low frequency interference usually leads to wandering or rough baseline. The surgery filter reduces
artifacts and interference from electrosurgical units. Under normal measurement conditions,
selecting
Surgery
may suppress certain features or details of the QRS complexes.
◆
ST
: recommended for ST monitoring.
10.6.4.7
Switching On or Off the Notch Filter
The notch filter removes the line frequency interference. To switch on or off the notch filter, follow this
procedure:
1.
Select the ECG numeric area or waveform area to enter the
ECG
menu.
2.
Select the
Setup
tab.
3.
Switch on or off
Notch Filter
.
NOTE
•
The notch filter can only be switched on or off when ECG Filter is set to Diagnostic. In other filter
modes, the notch filter is always on.
10.6.5
Disabling the Smart Lead Off Function
The monitor provides the smart lead off function . When the lead of the first ECG wave is detached but another
lead is available, the monitor automatically switches to the available lead to recalculate heart rate, and to analyze
and detect arrhythmias. When you reconnect the detached leads, the monitor automatically switches back to
the original lead.
The smart lead off function is enabled by default. To disable this function, follow this procedure:
1.
Select the ECG numeric area or waveform area to enter the
ECG
menu.
2.
Select the
Setup
tab.
3.
Switch off
Smart Lead
.
10.6.6
Disabling the CrozFusion
TM
Function
The CrozFusion
TM
function is enabled by default. However, in some situations you may need to disable this
function, or the CrozFusion
TM
function may not be able to work. You shall disable the CrozFusion
TM
function in
the following situation:
■
Administrating CPR
■
Performing CPB
■
Administrating IABP
■
Other situations that the CrozFusion
TM
function is not applicable
To disable the CrozFusion
TM
function, follow this procedure:
1.
Select the ECG numeric area or waveform area to enter the ECG menu.
2.
Select the
Setup
tab.
3.
Switch off
CrozFusion
.
WARNING
•
The monitor is used for single patient at a time. Simultaneously monitoring more than one patient
may result in a hazard to the patient.
•
ECG signal and Pleth signal from different patients may result in incorrect signal fusion.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...