NOTE
•
The Auto Limits for New Patient function is intended for newly admitted patients only.
•
The automatically set alarm limits take effect only after being confirmed.
9.6.4
Initiating Auto Alarm Limits
The monitor provides the auto alarm limits function to automatically adjust alarm limits according to the
patient’s vital signs using. When auto limits are selected, the monitor calculates safe auto limits based on the
latest measured values. To get accurate auto alarm limits, you need to collect a set of measured vital signs as a
baseline.
To initiate auto alarm limits, follow this procedure:
1.
Access the
Limits
page in either of the following ways:
◆
Select the
Alarm Setup
quick key
→
select the
Limits
tab.
◆
Select the
Main Menu
quick key
→
from the
Alarm
column select
Limits.
2.
From the
Limits
page, select
Auto Limits
at the left bottom.
3.
Select
OK
from the popup dialog box.
Then the monitor will automatically calculate alarm limits basing on the latest measured values. Before applying
these automatically created alarm limits, confirm if they are appropriate for your patient from the
Limits
menu. If
not, you can adjust them manually. These alarm limits will remain unchanged until you select auto limits again
or adjust them manually.
The monitor calculates auto limits basing on the following rules:
Module
Parameter
Lower Limit
Upper Limit
Auto Limit Range
Adult/
Pediatric
Neonate
Adult/
Pediatric
Neonate
ECG
HR/PR
(bpm)
HR × 0.8 or 40
(whichever is
greater)
(HR - 30) or 90
(whichever is
greater)
HR × 1.25 or
240
(whichever is
smaller)
(HR + 40) or
200
(whichever is
smaller)
Adult/Pediatric: 35 to 240
Neonate: 55 to 225
Resp
RR (rpm)
RR × 0.5 or 6
(whichever is
greater)
(RR - 10) or 30
(whichever is
greater)
(RR × 1.5) or
30 (whichever
is smaller)
(RR + 25) or
85 (whichever
is smaller)
Adult/Pediatric: 6 to 55
Neonate: 10 to 90
SpO
2
SpO
2
(%)
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
NIBP
NIBP-S
(mmHg)
(SYS × 0.68 +
10)
(SYS - 15) or
45 (whichever
is greater)
(SYS × 0.86 +
38)
(SYS + 15) or
105
(whichever is
smaller)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 35 to 115
NIBP-D
(mmHg)
(Dia × 0.68 +
6)
(Dia - 15) or
20 (whichever
is greater)
(Dia × 0.86 +
32)
(Dia + 15) or
80 (whichever
is smaller)
Adult: 25 to 225
Pediatric: 25 to 150
Neonate: 20 to 90
NIBP-M
(mmHg)
(Mean × 0.68
+ 8)
(Mean - 15) or
35 (whichever
is greater)
(Mean × 0.86
+ 35)
(Mean + 15 or
95)
(whichever is
smaller)
Adult: 30 to 245
Pediatric: 30 to 180
Neonate: 25 to105
Temp
(xx refers
to
temperat
ure site)
Txx (°C)
(Txx - 0.5)
(Txx - 0.5)
(Txx + 0.5)
(Txx + 0.5)
1 to 49
TD (°C)
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...