10 - 22
To set ST point, ISO point, and J point, follow this procedure:
1.
Select the ST numeric area, ECG numeric area, or ECG waveform area to enter the
ECG
menu.
2.
Select the
ST
tab
→
select the
Adjust
tab.
3.
Set
ST Point
.
The setting of
Auto Adjust
defines the method of adjusting the ISO point and J point.
Auto Adjust
is enabled by
default. In this case, positions of ISO point and J point are automatically adjusted accordingly. If you disable
when
Auto Adjust
, you need to manually adjust the position of ISO point and J point by selecting the arrows at
the right sides of
ISO
and
J
.
■
The ISO point (isoelectric) position is given relative to the R-wave peak. Position the ISO point in the middle
of the flattest part of the baseline (between the P and Q waves).
■
The J point position is given relative to the R-wave peak and helps locating the ST point. Position the J point
at the end of the QRS complex and the beginning of the ST segment.
■
The ST point is positioned a fixed distance from the J point. Move the J point to position the ST point at the
midpoint of the ST segment. Position the ST point relative to the J point at
J+60/80ms
,
J+40ms
,
J+60ms
or
J+80ms
. When
J+60/80ms
is selected, the ST point will be positioned 80 ms (heart rate 120 bpm or less) or
60 ms (heart rate more than 120 bpm) from the J point.
10.9
QT/QTc Interval Monitoring
The QT interval is defined as the time between the beginning of the Q-wave and the end of the T-wave. It
measures the total duration of ventricular depolarization (QRS duration) and repolarization (ST-T). QT interval
monitoring can assist in the detection of long QT syndrome.
The QT interval has an inverse relationship to heart rate. Faster heart rates shorten the QT interval and slower
heart rates prolong the QT interval. Therefore, several formulas can be used to correct the QT interval for heart
rate. The heart rate corrected QT interval is abbreviated as QTc.
QT/QTc interval monitoring is intended for adult, pediatric and neonatal patients.
10.9.1
QT/QTc Monitoring Limitations
Some conditions may make it difficult to achieve reliable QT/QTc monitoring, for example:
■
R-wave amplitudes are too low
■
The presence of frequent ventricular ectopic beats
■
Unstable RR intervals
■
P-waves tending to encroach on the end of the previous T-wave at high heart rates
■
The T-wave is very flat or T-wave are not well defined
■
The end of the T-wave is difficult to delineate because of the presence of U-waves
■
QTc measurements are not stable
■
In the presence of noise, asystole, ventricular fibrillation, atrial fibrillation, and ECG lead off
For these cases you should select a lead with good T-wave amplitude and no visible flutter activity, and without
a predominant U-wave or P-wave.
Some conditions such as left or right bundle branch block or hypertrophy can lead to a widened QRS complex. If
a long QTc is observed you should verify it to ensure that it is not caused by QRS widening.
Because normal beats followed by ventricular beats are not included in the analysis, no QT measurement will be
generated in the presence of a bigeminy rhythm.
If the heart rate is extremely high (over 150bpm for adults and over 180bpm for pediatrics and neonates), QT will
not be measured. When the heart rate changes, it can take several minutes for the QT interval to stabilize. For
reliable QTc calculation it is important to avoid measurements when the heart rate is changing.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...