9 - 7
9.6.5
Setting the Alarm Delay Time
For continuously measured parameters, you can set the alarm delay time. If the alarm condition is resolved
within the delay time, the monitor does not present the alarm.
This setting is password protected. For more information, see
.
The setting of
Alarm Delay
is not applied to the apnea alarms and the ST alarms. You can set
Apnea Delay
and
ST Alarm Delay
separately.
WARNING
•
The alarm delay time can be set to a maximum of 15 seconds. Changing this setting to an
inappropriate level could result in a hazard to the patient.
9.6.5.1
Setting the Apnea Delay Time
To set the apnea delay time, follow this procedure:
1.
Access the
Setup
page in either of the following ways:
◆
Select the
Alarm Setup
quick key
→
select the
Setup
tab.
◆
Select the
Main Menu
quick key
→
from the
Alarm
column select
Setup.
2.
Select
Apnea Delay
to set the apnea delay time.
9.6.6
Restoring the Default Alarm Settings
To reset all alarm settings to the defaults, follow this procedure:
1.
Access the
Alarm
page in either of the following ways:
◆
Select the
Alarm Setup
quick key.
◆
Select the
Main Menu
quick key
→
from the
Alarm
column select
Limits.
2.
On the
Limits
page, select
Defaults
at the bottom.
AG
EtCO
2
Same as the CO
2
module
FiCO
2
Same as the CO
2
module
awRR
(rmp)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
awRR+25 or
85 (whichever
is smaller)
Adult/Pediatric: 6 to 55
Neonate: 10 to 90
FiAA/EtAA
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
FiO
2
/EtCO
2
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
FiN
2
O/
EtN
2
O
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
C.O.
TB (°C)
Adult: (TB - 1)
Pediatric: N/A
N/A
Adult: (TB + 1)
Pediatric: N/A
N/A
Same as the
measurement range
C.I.
N/A
N/A
N/A
N/A
N/A
TFC
N/A
N/A
N/A
N/A
N/A
Module
Parameter
Lower Limit
Upper Limit
Auto Limit Range
Adult/
Pediatric
Neonate
Adult/
Pediatric
Neonate
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...