27 - 7
27.5
The CAA Settings
27.5.1
The EWS Tab
Other
Lethal Arrhy Alarms
Off
Disable
Selects whether lethal arrhythmia alarms can be
switched off.
•
Disable
: lethal arrhythmia alarms cannot be
switched off.
•
Enable
: lethal arrhythmia alarms can be
switched off from the
ECG
menu.
SpO2 Desat Alarm
Off
Disable
Selects whether the SpO
2
Desat alarm can be
switched off.
•
Disable
: the SpO
2
Desat alarm cannot be
switched off.
•
Enable
: the SpO
2
Desat alarm can be
switched off.
Apnea Alarm Off
Disable
Selects whether the apnea alarm can be switched
off.
•
Disable
: the apnea alarm cannot be switched
off.
•
Enable
: the apnea alarm can be switched off.
Arrhy Shield Time
2 min
Alarm light and alarm tone will be disabled for
designated period of time when certain
arrhythmia alarms are detected.
0
: disables this function.
Intubation Mode
Period
2 min
Selects the time for intubation.
Print on Alarm
Printer
Printer
: enables automatic printing via printer
when a parameter alarm is triggered.
Recorder
: enables automatic recording via
recorder when a parameter alarm is triggered.
CMS/eGW
Disconnected
Alarm
Off
Selects whether an alarm is issued when the
monitor is not connected or disconnected from
the CMS/eGateway.
Off
: the “Offline” alarm is not presented when the
monitor is not connected or disconnected from
the CMS/eGateway.
Section
Menu Item
Default Setting
Description
Menu Item
Default Setting
Description
Clinician ID
Off
Selects whether to allow inputting the clinician
ID to associate with the EWS score.
Clinician ID Timeout
10 min
Selects how long the clinician ID will remain valid
Default Adult Score
NEWS
Selects the default scoring tool for different
patient categories
Default Ped Score
/
Default Neo Score
/
Manage Score
Local
/
Delete
: deletes the selected scoring tools.
The monitor provide MEWS,NEWS and NEWS2 by
default. You cannot delete them.
USB Drive
/
Import
: imports the desired scoring tools to the
monitor.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...