D - 7
D.2.8
C.O. Technical Alarm Messages
D.2.9
CO
2
Technical Alarm Messages
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
TB Sensor Off
Low
A
Check the sensor connection and reconnect the
sensor.
TI Sensor Off
Low
A
Check the sensor connection and reconnect the
sensor.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
CO2 Module High Temp
Low
C
Ambient temperature is too high or there is a
module failure.
1. Lower the operating temperature.
2. Replug the module.
3. If the alarm persists, the CO
2
module may fail,
contact your service personnel.
CO2 Module Low Temp
Low
C
Ambient temperature is too low or there is a
module failure.
1. Raise the operating temperature.
2. Replug the module.
3. If the alarm persists, the CO
2
module may fail,
contact your service personnel.
CO2 Zero Failed
Low
C
For mainstream CO
2
module, check the
connections between the adapter and CO
2
transducer. Wait till the sensor’s temperature
becomes stabilized, and then perform a zero
calibration again.
For sidestream CO
2
module, replug the module. If
the alarm persists, contact your service
personnel.
CO2 No Watertrap
Low
B
Check the watertrap connections.
CO2 High Airway Press.
Low
C
1. Check the airway pressure settings of the
ventilator/anesthesia machine.
2. Disconnect the module from the ventilator/
anesthesia machine.
3. Replug the module.
4. If the alarm persists, contact your service
personnel.
CO2 Low Airway Press.
Low
C
1. Check the airway pressure settings of the
ventilator/anesthesia machine.
2. Disconnect the module from the ventilator/
anesthesia machine.
3. Replug the module.
4. If the alarm persists, contact your service
personnel.
CO2 High Barometric
Low
C
The ambient pressure exceeds the operating
pressure range or CO
2
module fails.
1. Make sure that the ambient pressure meets the
specifications, and check for sources that affect
the ambient pressure.
2. Replug the module.
3. If the alarm persists, contact your service
personnel.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...