22 - 4
◆
Scale 2
: f
or patients with a prescribed oxygen saturation requirement of 88–92% (for example, in
patients with hypercapnic respiratory failure).
3.
Measure or manually enter other required parameters and observations.
4.
If
the clinician ID is
enabled, input the clinician information by selecting
, and then
manually entering the information, or by scanning the clinician’s barcode.
5.
Select
Calculate
to get the total score.
6.
If
Score Confirmation
is enabled, select
Confirm
to save current scoring, or select
Cancel
to give up
current scoring. Refer to section
22.1.5.2 Setting the Scoring Confirmation Switch
for more information.
CAUTION
•
The decision to use Scale 2 of the SpO2 Scale should be made by a competent clinical decision maker
and should be recorded in the patient’s clinical notes.
NOTE
•
Before calculating the score, select Reset to clear the previous score.
•
The keyboard symbol at the right of the parameter value indicates that the value is manually
entered.
•
You can get the score only when all required parameters have been measured or entered.
•
When a patient is discharged or the monitor is turned off, the clinician ID is cleared.
22.1.4
EWS Alarm
If enabled, the monitor can automatically give alarms and refreshes the score.
22.1.4.1
Setting the EWS Alarm
If enabled, the monitor can automatically give alarms in the following cases:
■
The total score exceeds the configured threshold
■
The score of auto obtained parameter is 3.
To configure the EWS alarm, follow this procedure:
1.
From the EWS page select
Setup
.
2.
Select the
Alarm
tab.
3.
Turn on the
Alarm
switch.
4.
Set the alarm switches for the single parameters listed in the
3 in single parameter
area.
5.
Set the alarm switch and threshold of the total score in the
EWS Score
area.
22.1.4.2
Auto Refreshing Scores
If enabled, the monitor can automatically refresh the score in the following cases:
■
The total score reaches the configured threshold, or falls from the configured threshold to a lower score.
■
The score of auto obtained parameter reaches 3, or falls from 3 to a lower score.
To enable the auto refreshing score function, follow this procedure:
1.
From the EWS page select
Setup
.
2.
Select the
Alarm
tab.
3.
Turn on the
Auto Refresh Scores
switch.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...