A - 21
Microstream CO
2
Module
N
2
O
≤60
±1 mmHg
Hal
≤4
Sev
≤5
Iso
≤5
Enf
≤5
Des
≤15
±2 mmHg
*: means an extra error should be added in case of gas interference when CO
2
measurements are performed between 0 to
40mmHg.
Effect of interference gases on O
2
measurements
Gas
Quantitative effect
CO
2
0.2%
N
2
O
0.2%
Hal, Des, Sev, Iso, Enf
1%
Standard
Meet the standard of ISO 80601-2-55
CO
2
Measurement range
0 to 150 mmHg
Accuracy*
0 to 38 mmHg:
39 to 150 mmHg:
±2 mmHg
±5% of the reading (0.08% increased in error for
every 1 mmHg if the reading is more than 38
mmHg)
Accuracy drift
Meet the requirement for measurement accuracy within 6 hours
* Accuracy applies for respiration rate up to 80 rpm. For respiration rate above 80 rpm and EtCO
2
exceeding 18 mmHg, the
accuracy is 4 mmHg or ±12% of the reading, whichever is greater. For respiration rate above 60 rpm, the above accuracy
can be achieved by using the FilterLine H Set for Infant/Neonatal (Model: 006324). In the presence of interfering gases, the
above accuracy is maintained to within 4%.
Resolution
1 mmHg
Recovery time
<15 s (after defibrillation)
Sample flow rate
ml/min
Initialization time
30 s (typical)
180 s (maximum)
Response time
2.9 s (typical)
(The response time is the sum of the rise time and the delay time when
using a FilterLine of standard length)
Rise time: <190 ms (10% to 90%)
Delay time: 2.7 s (typical)
awRR measurement range
0 to 150 rpm
awRR measurement accuracy
0 to 70 rpm:
71 to 120 rpm:
121 to 150 rpm:
±1 rpm
±2 rpm
±3 rpm
awRR resolution
1 rpm
Data sample rate
40 Hz
50
+15
-7.5
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...