7 - 5
7.5.2.1
About Alarm Watch Tile
The main screen can display up to three alarm watch tiles, namely A-W1 and A-W2. Each tile can accommodate
up to six beds.
The following figure shows the alarm watch tiles.
The alarm watch tile is similar to alarm watch area in the
Remote View.
For more information, see
.
7.5.2.2
Displaying the Alarm Watch Tile on the Main Screen
To configure the alarm watch tile to be displayed on the monitor’s main screen, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Display
column select
Choose Screen
to enter the
Screen
Setup
menu.
2.
Select the
Tile Layout
tab.
3.
Select the numeric area where you want to display the alarm watch tile, and then in the drop-down list,
select
Alarm Watch
→
A-W1
or
A-W2
.
7.5.3
Auto Displaying the New Alarm Bed
The monitor provides the function of automatically displaying the remote alarm bed. If this function is enabled,
when a remote bed issues an alarm, the monitor automatically displays the monitoring information from this
remote bed.
If multiple remote beds issue alarms, the monitor cyclically displays the alarm beds as per the preset interval and
in the order of alarm time.
The auto displaying alarm bed function is disabled by default. To enable this function, follow this procedure:
1.
From the
Remote View
screen, select
to enter the
Remote View
setup menu.
2.
Switch on
Rollup Alarm Beds
.
3.
Set
Rollup Interval
:
◆
Off
: do not cyclically display the remote alarm beds. Once a new alarm is issued, the monitor
automatically switches to the new alarm bed.
◆
10 sec
,
20 sec
, or
30 sec
: If multiple remote beds issue alarms, the monitor cyclically displays the
alarm beds as per the preset interval and alarm priority in the order of alarm time.
4.
Set
Alarm Priority
:
(1)
(5)
(4)
(3)
(2)
(1)
(3)
(1)
Alarm watch tile label
(2)
Disconnection icon: when the remote device is disconnected, this icon displays at the tile, and the tile
background color is red.
(3)
Select bed icon: select it to enter the
Select Bed
window.
(4)
More than one bed tile: when more than one bed is assigned to a tile, the tile displays the alarm status,
connection status, etc.
(5)
One bed tile: when only one bed is assigned to a tile, the tile displays the parameter value and alarm
message from this bed, etc.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...