20 - 7
•
Switching the auto impedance check off will disable automatic prompt to the user of impedance
value changes, which may lead to incorrect BIS values. Therefore, this should only be done if the
check interferes with or disturbs other measurements.
20.8.3
Manual Sensor Check
To manually start a sensor check, use either of the following method:
■
Press the sensor check hardkey
on the BIS module.
■
Select the
Sensor Check
tab from the
BIS
menu.
The monitor enters the sensor check window after you start the sensor check. The sensor check window displays
the following items:
■
Sensor Type
■
The status of each electrode
■
Expiration time or usable cycles
NOTE
•
For different types of sensors the sensor check window may be different.
20.8.4
BIS Sensor Status
The color of each electrodes indicates its status:
Although BIS may still be measured when the electrode status is
Noise
or
High
, for best performance, all
electrodes should be in
Pass
status.
The sensor check may fail for the following reasons:
■
Impedance too high
■
Incorrect sensor application
■
Poor sensor connection
■
Defective patient interface cable or sensor
To correct the situation:
■
Recheck the sensor
■
Reapply the sensor according to instructions
■
Check sensor connection
■
Replace patient interface cable or sensor
20.9
Monitoring Bilateral BIS
By using BISx4 and the Bilateral sensor you can perform bilateral BIS monitoring. You can display the BIS expand
view during bilateral BIS monitoring.
Color
Status
Description
Action
Red
Lead off
Electrode falls off and has no skin
contact
Press the sensor more firmly to skin to ensure
good sensor-to-skin contact. If necessary,
remove the sensor, and then clean and dry the
skin. Reapply the senor or replace the sensor.
Grey
Noise
The EEG signal is too noisy. Impedance
cannot be measured
Press the sensor more firmly to skin to ensure
good sensor-to-skin contact.
Yellow
High
The impedance is above the limit
Press the sensor more firmly to skin to ensure
good sensor-to-skin contact.
Green
Pass
The impedance is within valid range
No action necessary.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...