28 - 4
28.4.3
Battery-related Alarms
The capacity of the battery is limited. When the battery is low, the monitor presents the
Low Battery
alarm, the
alarm lamp flashes, and the monitor produces an alarm sound.
If the battery is almost depleted, the monitor presents the
Critically Low Battery
alarm. In this case,
immediately connect the AC mains to power the monitor and charge the battery. Otherwise, the monitor will
automatically shut down soon.
For more information on battery-related alarms, see
28.5
Charging the Battery
The battery is recharged automatically when the monitor is connected to the external power supply.
28.6
Maintaining the Battery
28.6.1
Conditioning the Battery
The performance of batteries deteriorates over time. You should condition the batteries every three months.
If the battery is not conditioned for a prolonged time, its charge indication may not be accurate and you may
wrongly evaluate the remaining battery runtime.
To condition a battery, follow this procedure:
1.
Disconnect the equipment from the patient and stop all monitoring and measuring procedures.
2.
Allow the battery to be charged uninterruptedly till it is fully charged.
3.
Allow the equipment to run on the battery until the battery is completely depleted and the equipment
automatically shuts down.
4.
Fully charge the battery again for use or charge it to 40 – 60% for storage.
NOTE
•
Do not use the equipment to monitor the patient during battery conditioning.
•
Do not interrupt battery conditioning.
28.6.2
Checking Battery Performance
The performance of a rechargeable battery deteriorates over time. You should check the battery performance
every three months or if you doubt that the battery may fail.
See steps 1 to 3 of
28.6.1 Conditioning the Battery
to check battery performance. The operating time of the
batteries reflects their performance directly. If the operating time of a battery is noticeably shorter than that
stated in the specifications, the battery may reach its service life or malfunction.If the battery performance
meets the requirement, fully charge the battery again for use or charge it to 40 – 60% for storage.
NOTE
•
Battery operating time depends on equipment configuration and operation. For example, high
display brightness or measuring NIBP repeatedly will shorten the battery operating time.
28.7
Storing Batteries
When storing batteries, make sure that the battery terminals do not come into contact with metallic objects. If
batteries are stored for an extended period of time, place the batteries in a cool place with a partial charge of
40% to 60% capacity.
Condition the stored batteries every three months. For more information, see
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...