16 - 5
16.6.2
Changing the Pressure Label
The pressure label is a unique identifier for each type of pressure. Therefore, you should select a proper pressure
label for the source of the pressure you want to monitor.
To select the pressure label, follow this procedure:
1.
Select the IBP numeric area or waveform area to enter the corresponding pressure menu.
2.
Select the
Setup
tab.
3.
Set
IBP1 Label
or
IBP2 Label
.
NOTE
•
It is not allowed to select the same label for different pressures.
16.6.3
Setting the Pressure Type for Display
For the non-specific pressure (P1, P2, P3 or P4), the displayed pressure type is configurable. To set the displayed
pressure type, follow this procedure:
1.
Select the numeric area or waveform area of the non-specific pressure to enter the corresponding pressure
menu.
2.
Select the
Setup
tab.
3.
Set
Measure
:
◆
If this non-specific pressure is artery pressure, set the
Measure
to
All
. In this case, its corresponding
numeric area displays systolic pressure, diastolic pressure and mean pressure.
◆
If this non-specific pressure is venous pressure, set the
Measure
to
Mean Only
. In this case, its
corresponding numeric area displays only the mean pressure.
16.6.4
Changing the Sensitivity
The IBP value displayed on the monitor screen is the average of data collected within a specific time. The shorter
the averaging time is, the quicker the monitor responds to changes in the patient’s blood pressure, and the
higher the sensitivity. Contrarily, the longer the averaging time is, the slower the monitor responds to changes in
the patient’s blood pressure, the lower the sensitivity, but the measurement accuracy will be improved. For
critically ill patients, selecting higher sensitivity will help understanding the patient’s state.
To set the sensitivity, follow this procedure:
1.
Select the IBP numeric area or waveform area to enter the corresponding pressure menu.
2.
Select the
Setup
tab.
3.
Set
Sensitivity
.
Label
Description
Label
Description
PA
Pulmonary artery pressure
CVP
Central venous pressure
Ao
Aortic pressure
LAP
Left atrial pressure
UAP
Umbilical arterial pressure
RAP
Right atrial pressure
BAP
Brachial arterial pressure
ICP
Intracranial pressure
FAP
Femoral arterial pressure
UVP
Umbilical venous pressure
Art
Arterial blood pressure
LV
Left ventricular pressure
CPP
Cerebral perfusion pressure
P1 to P4
Non-specific pressure label
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...