E - 2
E.3
Device Labeling
Check the labels provided by the manufacturer or the healthcare facilities are present and legible.
■
Main unit label
■
Integrated warning labels
E.4
Protective Earth Resistance
1.
Plug the probes of the analyzer into the device’s protective earth terminal and protective earth terminal of
the AC power cord.
2.
Test the earth resistance with a current of 25 A.
3.
Verify the resistance is less than limits.
LIMITS
For all countries, R = 0.2 Ω Maximum
E.5
Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage tests.
The following outlet conditions apply when performing the Earth Leakage test:
■
normal polarity (Normal Condition),
■
reverse polarity (Normal Condition),
■
normal polarity with open neutral (Single Fault Condition),
■
reverse polarity with open neutral (Single Fault Condition)
LIMITS
For UL60601-1,
◆
300
μ
A in Normal Condition
◆
1000
μ
A in Single Fault Condition
For IEC60601-1,
◆
500
μ
A in Normal Condition
◆
1000
μ
A in Single Fault Condition
E.6
Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All measurements have
a true RMS only
The following outlet conditions apply when performing the Patient Leakage Current test.
■
normal polarity (Normal Condition);
■
reverse polarity (Normal Condition),
■
normal polarity with open neutral (Single Fault Condition);
■
reverse polarity with open neutral (Single Fault Condition).
■
normal polarity with open earth (Single Fault Condition);
■
reverse polarity with open earth (Single Fault Condition).
LIMITS
For CF
applied parts
◆
10 μA in Normal Condition
◆
50 μA in Single Fault Condition
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...