17 - 6
Problem
Solution
Do not see C.O. numeric area on the main
screen
1. Check that the C.O. is set to display in the
Screen Setup
menu. For
more information, see
.
2. Check that if the C.O. parameter switch is enabled. If not, enable the
C.O. measurement. For more information, see
.
3.Check that the patient type is adult.
4. Check the connection of C.O. cable, thermodilution catheter and TI
sensor.
C.O. value is inaccurate
1. Check that the thermodilution catheter is positioned properly.
2. Check that the computational constant is proper for current injectate
temperature, injectate volume and injectate probe type.
3. Inject solution rapidly and smoothly.
4. Finish injection within four to five seconds.
5. Inject more volume, or inject colder solution.
6. Check that the height and weight of patient is properly configured.
7. If
Auto TI
is switched off, check that the entered temperature is
correct.
C.O. measurement fails
1. Inject more volume, or inject colder solution. Make sure that the
injectate temperature is at least 10°C colder than the patient blood
temperature.
2. Finish injection within four to five seconds.
3. Check the connection of C.O. cable, thermodilution catheter and TI
sensor.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...