A - 10
Pace pulse markers
Pace pulses meeting the following conditions are labelled with a PACE
marker:
Amplitude:
Width:
Rise time:
No overshoot
±2 to ±700 mV
0.1 to 2 ms
10 to 100 μs (no greater than 10% of pulse width)
Pace pulse rejection
When tested in accordance with the IEC 60601-2-27: 201.12.1.101.13, the
heart rate meter rejects all pulses meeting the following conditions.
Amplitude:
Width:
Rise time:
No overshoot
±2 to ±700 mV
0.1 to 2 ms
10 to 100 μs (no greater than 10% of pulse width)
HR
Measurement range
Neonate:
Pediatric:
Adult:
15 to 350 bpm
15 to 350 bpm
15 to 300 bpm
Resolution
1 bpm
Accuracy
±1 bpm or ±1%, whichever is greater.
Sensitivity
200 μV (lead II)
HR averaging method
In compliance with the requirements in Clause 201.7.9.2.9.101 b) 3) of IEC
60601-2-27, the following method is used:
If the last 3 consecutive RR intervals are greater than 1200 ms, the 4 most
recent RR intervals are averaged to compute the HR. Otherwise, heart rate is
computed by subtracting the maximum and minimum ones from the most
recent 12 RR intervals and then averaging them.
The HR value displayed on the monitor screen is updated no more than one
second.
Response to irregular rhythm
In compliance with the requirements in Clause 201.7.9.2.9.101 b) 4) of IEC
60601-2-27, the heart rate after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy (waveform A1): 80±1 bpm
Slow alternating ventricular bigeminy (waveform A2): 60±1 bpm
Rapid alternating ventricular bigeminy (waveform A3): 120±1 bpm
Bidirectional systoles (waveform A4): 90±2 bpm
Response time to heart rate change
Meets the requirements of IEC 60601-2-27: Clause 201.7.9.2.9.101 b) 5).
From 80 to 120 bpm:
From 80 to 40 bpm:
less than 11 s
less than 11 s
Time to alarm for tachycardia
Meets the requirements in Clause 201.7.9.2.9.101 b) 6) of IEC 60601-2-27.
Waveform
B1h-range:
B1-range:
B1d-range:
B2h-range:
B2-range:
B2d-range:
<11 s
<11 s
<11 s
<11 s
<11 s
<11 s
Tall T-wave rejection capability
When the test is performed based on Clause 201.12.1.101.17 of IEC 60601-2-
27, the heart rate calculation is not affected for QRS of 1 mV amplitude and
100 ms
duration, T-wave duration of 180 ms and amplitude lower than 1.2
mV, and QT interval of 350 ms.
Arrhythmia Analysis Classifications
Asystole, V-Fib/V-Tac
h
, V-
T
ac
h
, Vent-. Brady, Extreme Tachy, Extreme Brady,
Vent Rhythm, PVCs/min, Pauses/min, Couplet, Bigeminy, Trigeminy, R on T,
Run PVCs, PVC, Tachy, Brady, Missed Beat, Pacer Not Paceing, Pacer Not
Capture, Multiform PVC, Nonsus. V-Tac, Pause, Irr. Rhythm, A-Fib
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...