6 - 8
ADM 6725 E10
KA
T
Program Branch OUTPUT OPTIONS
FLUXUS can generate the following columns:
Column title
Column format
Contents
MEASURE
000000.00 the quantity of measurement which was in >Output Options< selected
Q_POS
+00000000.00 total of measuring quantity in positive (forward) flow direction
Q_NEG
-00000000.00 total of measuring quantity in negative (reverse) flow direction
T1 000.0
temperature
T1
T2 000.0
temperature
T2
...
Other available temperature inputs or process inputs
Columns with ONLINE serial output (during measurement)
In ONLINE mode, columns will be generated for all quantities which
may
be output during
measurement.
(The columns Q_POS and Q_NEG can remain empty, in case the totaliser function has not been
activated.)
Since for the measuring quantity ‘flow velocity’ no total counting can be enabled,
no columns for total values will be generated.
Columns with OFFLINE serial output (stored measured values)
In OFFLINE mode, columns will
only
be generated if at least one measured value was
stored in the respective data set.
(The columns Q_POS and Q_NEG are not generated if the totaliser function was not enabled.)
6.7 Set-up of the Storage Interval
A storage interval or rate must be set for:
•
the internal data logger of the instrument,
•
serial output ONLINE,
The set storage rate will be active for the above-mentioned options.
Note:
•
If
none
of these options are activated, FLUXUS will omit the display group
'Storage Rate'.
•
In
multi-channel measurement
mode, the storage interval can be set for each
channel independently. The storage interval in seconds for any measuring chan-
nel should be at least equal to the number of activated measuring channels.
(When 2 measuring channels are activated, the storage rate for a channel should be at least 2 seconds,
not less.)
This is the display for setting the storage interval:
Storage Rate
"
Once per 10 sec.
O11-1
Scroll through this display group to set
the desired storage interval
(in multi-channel
measurement for the currently selected measuring
channel)
.
#
Every
Minute
O11-2
#
Once per 10 Min.
O11-3
#
Once per 30 Min.
O11-4
#
Every
Hour
O11-5
$
EXTRA
O11-6
#
Every
Second
O11-7
To
current
loop, e.g.
(Without pressing ENTER, the display
list scrolls around)
Confirm the set storage interval by
pressing
ENTER.
Storage Rate
"
EXTRA
O11-6
If you select
EXTRA
, you may set a
storage interval in the range from 1 to
43200 seconds (12 hours)
as per
requirement
of the application.
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...