3 - 2
ADM 6725 E10
KA
T
Selection of the Measuring Point
Select the measuring point so that the sound waves from the transducers
propagate the pipe horizontally.
(Solid particles are deposited on the bottom of the pipe, gas pockets can develop at the top.)
Correct
Incorrect
Free inlet or outlet pipe sections
Select the measuring point at a location where the pipe cannot drain out.
Correct
Disadvantageous
Correct
Disadvantageous
Vertical pipe
Select the measuring point at a location where the liquid flows up.
(Pipe completely filled)
Correct
Incorrect
3.4 Distances between Measuring Point and
Disturbance Source
In the following examples, recommended
straight inlet and outlet pipe lengths are
given for different types of flow disturbance
sources to assist in selecting the correct
measuring point.
D
=
Nominal pipe diameter at measuring point
L = Recommended
distance
Disturbance source
: 90 °-elbow
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...