8 - 4
ADM 6725 E10
KA
T
Measuring Data Transfer to PC
/HD
-
Suppresses the output of column titles.
\DATA
7,13; +3,22; -0,44; 20,3; 40,8;
/XT+
With this option, nearly all data of the parameter records, even if they are not
significant for the particular measuring data set, will be output (e.g. measuring range
high limit value). Furthermore, the parameters used by FLUXUS for medium, pipe
and lining material as well as for the transducers will be output. In addition to the
sound path method, the number of sound paths is given.
/DT+
Measured values will be stamped with current data and time.
/PSPS
+
Only with multi-channel measurement: output of the parameter records next to each
other.
/PSPS
-
Only with multi-channel measurement: output of the parameter records before
respective measured values of that channel. (Default)
8.4 Utility Program MOVEDATA
MOVEDATA is a computer terminal program. With this software, your PC works in a
similar way to a serial printer.
Start the program MOVEDATA.EXE from the MS-DOS prompt using the following format:
SYNTAX:
movedata <OutputName> [<Options>]
PARAMETERS
:
<OutputName>
File name of the ASCII-file to be created. A file name must be
specified.
[<Options>]
/Cx
Specifies the serial port.
Without specifying this parameter,
COM2
will be used.
/C1=COM1 ... /C4=COM4
/N
An already existing file will be overwritten without
confirmation.
/H
Output of a help text.
Example:
movedata data1.txt /C1
FLUXUS will transmit the measured values via serial interface COM1. The binary file
‘data1.flx’ and ASCII-file ‘data1.txt’ will be created.
8.5 Program Information and Error Messages
(MOVEFAST/FLXtoTXT)
Screen Messages
Description
After start of program
(only applicable for MOVEFAST)
>> %s: THIS SERIAL PORT IS
NOT PRESENT ON THIS PC
The serial port is not available on the PC you are using. Since
MOVEFAST accesses COM2 by default, the program
parameter needs to be set to /C1 if you use a PC or Laptop with
only one serial interface, i.e. COM1.
>> INVALID /C PARAMETER
Only /C1, /C2, /C3, /C4 are accepted.
During file opening
(applicable for MOVEFAST and FLXtoTXT)
FILE '<name>' ALREADY EXIST !
OVERWRITE ? y/[N]
The file <name> already exists. It will be overwritten if you
enter ‘Y’ or ‘J’. The program will be terminated if you enter ‘N’.
This confirmation message will not appear if you start the
program with parameter /N.
Opening binary file '<name>'
FLXtoTXT is opening the file <name> that will be converted.
Creating binary file '<name>'
MOVEFAST is creating the binary file <name>.
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...