ADM 6725 E10
KA
T
4 - 9
Program Branch PARAMETER
4.3.4 Medium Temperature
FLUXUS uses the medium temperature for correcting the sound velocity and the
viscosity which both depend on temperature. However, the medium temperature
has no direct influence on the measuring result.
20 c
Medium Temperat.
P16 Enter the medium temperature within the
operating range of the transducer. The
default value is 20 °C.
Confirm your entry by pressing
ENTER
.
Note:
The medium temperature, which is accepted, depends on the operating range of the
selected transducers.
Measuring
par >MEA< opt sf
M1
When the medium temperature has been
entered, the program returns to the MAIN
MENU.
If FLUXUS cannot identify a transducer
connected to it, some other displays for
the sensor values will appear.
(See overview
in chapter 4.1)
4.3.5 Input of the Transducer Parameters
If FLUXUS cannot recognise a transducer connected to the transducer connection
port, the following displays will appear.
(The parameters for the standard transducer Type Q3N are
already pre-programmed.)
Standard
Transducer Type
"
P17
Here, confirm the transducer type.
(For example: Standard)
.
With the transducer type selected, FLUXUS ignores the rest of the display group
and jumps straight to the MAIN MENU again.
Using a HotCode, the low limit for the inner pipe diameter can be changed for the
transducer Type Q.
(See also chapter 2.4.4)
NEXT
M UX
Enter this key sequence from the MAIN
MENU.
25 mm
Dnmin Sensor’Q’
Here, the low limit for the inner diameter
for the transducer Type Q can be set,
which FLUXUS will accept.
This value can be within a range from 10
mm to 63 mm.
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...