7 - 4
ADM 6725 E10
KA
T
Program Branch SPECIAL FUNCTIONS
7.2
Working with Parameter Records
Parameter records are data that contain all necessary information to perform a
certain measurement task. These data are:
•
Pipe parameters
•
(Transducer
parameters)
•
Medium parameters
•
Output options
Parameter records can be created before each measurement and then be stored
via SPECIAL FUNCTIONS. FLUXUS can store up to 14 different parameter rec-
ords. If you wish to work with a stored parameter record, you need to select that
record as "Current Record". This way, repeat measurement tasks can be
performed with previously entered parameters much more easily and quickly.
7.2.1
Save Current Parameter Record
The parameters need to be entered first as previously described in chapters
"Program Branch PARAMETER" and "Program Branch OUTPUT OPTIONS". If
this has been done, the parameters can be stored as the current parameter
record.
Special Funct.
"
Store Curr.Rec.
S4
Use display S4 of program branch SF to
store the entered data as current
parameter record. Confirm by pressing
ENTER
.
Store Par. to:
"
Par.Record 01
S4-1
The display "Store Par. to:" appears and
offers a choice of 14 memory locations.
Confirm your selection by pressing
ENTER
.
Should you select and confirm S4-15
"Don't Store",
the current parameter
record will not be stored.
Par.Record 02
S4-2
...
...
After pressing ENTER, FLUXUS leaves
program branch SF and returns to the
MAIN MENU.
Par.Record 14
S4-14
Don't Store
S4-15
Overwrite
no >YES<
S4-16 Should the memory location be already
occupied by a stored parameter record,
FLUXUS asks to overwrite it.
!
Next memory
location
!
MAIN MENU
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...