14 - 4
ADM 6725 E10
KA
T
Errors
14.1.4 Error during Input of Programming Code
Error message F12:
INVALID CODE !
909049
Reason:
The programming code was not entered correctly.
Corrective action:
Check, whether
•
you have entered a reserved code (e.g. which is
used for selecting a language),
•
you have entered less than 6 characters.
Enter the programming code correctly.
14.1.5 Error Messages during Set-up of Date/Time
Error message F13:
Date 31.04.1998
INVALID DATE !
Reason:
The date was not entered correctly.
Corrective action:
Enter a valid date.
Error message F14:
Time 14:63
INVALID TIME !
Cause:
The time was not entered correctly.
Corrective action:
Enter a valid time.
14.1.6 System Errors
Error message F15:
SYSTEM ERROR !
207--0:7300
(Number sequence is an example.)
Reason:
Unexpected system error.
Corrective action:
COLD START the instrument by pressing these keys
at the same time.
4-56)46
INIT
BR K
E NT E R
.
Should this message appear several times, please
make a note of the serial number of your device and
of the number sequence in the lower line of the
display. Call your distributor for further assistance.
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...