5 - 1
Program Branch MEASURING
5 Program Branch MEASURING
The correct positioning of the transducers is very important and is a pre-condition for
error-free measurements. It guarantees that the sound signal will be received under
optimum conditions and evaluated correctly. Because of the variety of applications and
factors influencing measurement, there can be no standard solution for the positioning of
the transducers. The correct position of the transducers will be influenced by
•
pipe
diameter,
•
linings,
•
wall
thickness,
•
inner
deposits,
•
pipe
material,
•
medium.
5.1 Program Branch MEA Overview
The program branch
MEASURING
starts with the MAIN MENU and contains the following
displays:
Measuring
par>MEA< opt sf
M1
!
MEASUR
√
- . .
CHANN: >A<B Y Z
M3
Please select the channel for flow measurements (A, B
are flow channels; Y, Z are virtual calculation channels).
Tick sign ‘
√
’ = Channel is ON
Minus sign ‘-‘ = Channel is OFF
Dot sign ‘.’ = Channel not activated (parameters are not
set-up)
"
select
#
activate/deactivate
ENTER confirm
!
!
!
!
!
!
!
!
A:Meas.Point No.
M2
(This display will appear if in
program branch Output Opt-
ions 'Store Meas. Data', or
'Serial Output' has been
activated.
2 NUM
A:Sound Path
M4
The Letter A indicates that the measuring channel A is
activated.
The ‘Sound Path’ is the number of times the
ultrasonic signals transverses in the pipe.
(This
number can be edited, for further explanations see
chapter 5.2.2).
!
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...