ADM 6725 E10
KA
T
5 - 5
Program Branch MEASURING
Figure=number
of transits
Sound path
Figure=number
of transits
Sound path
1
2
3
4
5
6
etc. etc.
5 NUM
A:Sound Path
M4 Enter the required number of transits, in our
example 5.
Press ENTER to confirm.
A: 106 mm Diagon
Transd. Distance
M5 The calculated transducer distance is
displayed. With an odd number of transits,
´Diagon´ appears to show that the diagonal
mode is used.
(2) 5 NUM
A:Sound Path
M4* If the measurement was interrupted and the
Sound Path was changed, display M4 will also
indicate in brackets the number of transits last
used as a reminder that the positioning of the
transducers should be checked again.
5.2.3 Transducer Installation in Reflection Mode
2
A:Sound Path
(Letter A = Channel A)
M4
By entering an even number under Sound
Path, the reflection transducer mounting
method can be chosen.
Transducer 1 Transducer 2
In reflection mode, the transducers must be
mounted on the same side of the pipe wall.
The sound waves are reflected by the
opposite pipewall. Thus, the length of the
sound path in the pipe doubles. The
accuracy of the measurement increases.
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...