5 - 4
ADM 6725 E10
KA
T
Program Branch MEASURING
5.2.1 Identification of the Measuring Point
The measuring point can be identified
•
either by using plain-text (example: ´MS.PK20!´)
•
or by numerical values, decimal point, and/or slash (example: 18.05-06).
There are
eight
digits at your disposal to identify the measuring point.
In program branch SPECIAL FUNCTIONS, you can select which mode for the
identification of the measuring point will be used
(see chapter 7.7.2)
.
Additionally this set-up can be selected by using a HotCode entry, you can select the
respective mode and make it cold-start resistant.
(See also chapter 2.4.4)
NEXT
ISP
Immediately after pushing the key BRK, enter
this key sequence from the MAIN MENU.
>1234< ( )
Meas.Point No.:
↑↓← →
.
After entering above HotCode, this display
appears.
It is for selecting the input mode for the
identification of the measuring point.
Select and confirm the mode you prefer.
⇑
Identification of the
measuring point by
direct method through
numbers, point, slash.
⇑
Identification of
measuring point by
indirect method
through ASCII-editor
acc. to chapter 2.5
The following display for the identification of the measuring point will appear in program
branch MEASURING:
- ( )
Meas.Point No.:
↑↓← →
⇑
These symbols shall
be a reminder that the
ASCII input mode has
been selected.
M2
Should you have activated 'Store Measuring
Data' or 'Serial Output' in program branch
OUTPUT OPTIONS, you may enter here a
number for the identification of the
measuring point.
(A measuring point may or may not
carry an identification mark.)
Confirm the
identification mark for the measuring point,
and return to the program by pressing
ENTER
.
(Now, FLUXUS will mark all logged measuring
values by the entered identification.)
5.2.2 Decide which Sound Path to use
The Sound Path represents the number of transit paths that the ultrasonic waves travel
through the medium in the pipe.
Any
odd
number requires diagonal mounting of the transducers on different sides of the
pipe.
Any
even
number requires installation of the transducers so that the reflection method
can be used.
Synopsis
Transducer installation
in diagonal mode
Transducer installation
in reflection mode
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Page 4: ...ADM 6725 E10 KAT...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Page 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...