Chapter
5
Maintenance
5-50
Confidential
Copying Prohibited
●
Others
Item
Inspection Item
Procedure
Criteria
1 Clock
Setting
Press the
(Menu) key, and
touch [Initial Setting/Maintenance]
→
[Clock Setting], and check the
Clock Setting menu.
Current date/time should be
displayed.
2 Network
Setting
Press the
(Menu) key, and
touch [Initial Setting/Maintenance]
→
[Network and ID Reader
Setting], and check the Network
and ID Reader Setting menu.
Correct network should be
set.
3 Power Supply Cable
Perform continuity test using the
tester.
Should pass continuity test.
Others
4 Electric
Fan
Verify that electric fan is operating
when the power is turned ON.
Electric fan should be
properly operating.
5.9.4.2 Safety Inspection
Perform inspections according to the procedure explained in “5.9.5 Safety inspections”.
5.9.4.3 Cleaning and Disinfecting
Clean and disinfect according to the procedure explained in “5.2 Cleaning and Disinfecting”.
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...