Chapter 10 Part Location Diagram and Checkpoints
Confidential
Copying Prohibited
10-15
10.4 Power Supply Unit
10.4.1 Output Voltage Check
To check the output of the power supply only, temporarily set pin 19 to "L" (short it by connecting it
to GND) to run the power supply unit.
The output from each pin is as follows.
Pin No.
Output voltage
Allowable voltage
Cable color
Pin 1
+3.3V
+3.3
±
0.165V
Purple
Pin 2
+3.3V
+3.3
±
0.165V
Purple
Pin 5
+5V
+5.0
±
0.25V
Yellow
Pin 6
+5V
+5.0
±
0.25V
Yellow
Pin 9
+12V
+12.0
±
0.6V
Orange
Pin 10
+12V
+12.0
±
0.6V
Orange
Pin 13
+12V
+12.0
±
0.6V
Orange
Pin 15
+24V
+24.0
±
1.2V
Red
Pin 17
+24V
+24.0
±
1.2V
Red
Pin 22
+5V
+5.0
±
0.25V
Green