Chapter 8
Disassembly and Assembly
Confidential
Copying Prohibited
8-11
8.2 Instrument Assembly
8.2.1 Main
Block
Assembly
(1) Insert the air tube in the connector to assemble the main chassis SU.
(2) Connect the BP cable and air tube to the pump block assembly.
(3) Mount the pump block assembly parts on the main chassis SU in the following order: [Red]
Æ
[Yellow]
Æ
[Black]
Æ
[Green].
Place this below the main
chassis and screw it down.
Secure locations (2) marked
○
using:
W SEMS (L) M3 × 8
Red
Yellow
Black
Green
If the air tube is difficult to
insert, apply ethanol to the
connector.
[Red]
Insert section marked
○
in the
notch of the main chassis.
Make sure part is properly seated
before securing using screws!
[Yellow]
[Green]
Red
Yellow
Black
Green
BP cable connections
Insert air tube over
connector.
* Make sure that
surrounding cables do
not prevent full insertion
of tube.
Secure locations (4) marked using:
W SEMS (L) M3 x 8
[Black]
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...