5.9 Maintenance and Inspection
5.9.1 About Maintenance and Inspection
To maintain the safety and the reliable function and performance, daily inspection before use
and periodical inspection are indispensable. Fukuda Denshi is not responsible for troubles
caused by failure to observe the instructions described in this section.
CAUTION
•
Never disassemble or remodel the instrument.
It may result in fire hazard or electric shock.
•
Do not perform the Maintenance menu procedure which is not described in this manual.
It may result in malfunction.
•
Be careful not to have liquids such as alcohol or water enter the instrument.
•
Inspect the mainframe and all connectors and cables for any external damage.
If any of the following events occurs, immediately perform maintenance and inspection.
・
The instrument receives a strong impact due to dropping, etc.
・
Any liquid or flowing object falls on or inside the instrument.
・
The instrument does not function normally.
・
The enclosure of the instrument is cracked, broken or damaged.
・
Cables including the power cable and leadwires are deteriorated in performance.
●
Daily Check
Prepare a copy of the “Daily Check List” and perform the daily check everyday according to
the “Daily Check Procedure” below.
・
If the instrument fails any inspection item on the daily inspection list, the general
assessment must be "failed". Repair the instrument so that it passes all inspection items.
・
Use the instrument only when it has passed all inspection items.
●
Periodic Check
There are 2 periodic inspections that needs to be performed. “Periodic Inspection (1)" can
be performed without using any special jig or measurement device, and “Periodic Inspection
(2)" needs to use special jig and measurement device. Perform inspections according to the
procedure explained.
・
If the instrument fails any inspection on the periodic inspection list (1), (2), the overall
judgment will be "Fail". Repair the instrument so that it passes all inspection items.
・
Use the instrument only when it has passed all inspection items.
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...