7. TP4
Measures the voltage of supplementary pressure Q.
Use this voltage for adjusting the accuracy of supplementary pressure Q.
8. TP5
GND terminal.
9. Pressure pulse wave (OSC)
Check at pin 4 on CN2.
10. Cuff pulse wave (PW)
Check at pin 5 on CN2.
11. Pressure safety circuit 1 (LOWCUFF)
A 10 mmHg or higher pressure is output as "H".
Check at pin 18 on CN2.
12. Pressure safety circuit 2 (HICUFF)
A 320 mmHg or higher pressure is output as "H".
Check at pin 17 on CN2.