Chapter 8
Disassembly and Assembly
3. Connect the power cable (printer).
Connect the power cable from the printer AC
adapter to the printer outlet on the isolation
transformer.
Isolation Transformer
Printer AC adapter
Printer power cable
To printer outlet
4. Connect the cables (VS-1500, printer). For details refer to “Wiring Diagram”.
Connect the power cable (CS-40, accessory from the
printer shelf) from the VS-1500 to the isolation
transformer.
Connect the power cable from the isolation transformer
to a hospital grade 3-pronged outlet.
Isolation
Printer AC adapter
VS-1500
Printer
transformer
Power Cable
CS-40
Hospital grade
3-pronged outlet
5. Check the installation, wiring and connections of all devices again.
z
The printer shelf should slide smoothly.
z
No cables should be exposed to stress.
z
The printer power cable and the USB cable must be properly routed through the
cable guides to enable it to move with the sliding of the printer shelf.
z
You should be able to lock the printer shelf both in its retracted and extended
position.
8-32
Confidential
Copying Prohibited
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...