Chapter 12 Information of the electromagnetic compatibility (EMC)
Confidential
Copying Prohibited
12-3
●
Compatibility with Electromagnetic Immunity (1)
The VS-1500 is intended for use in the electromagnetic environments described below. It is
recommended to verify that the VS-1500 is used in such an environment.
Immunity Test
IEC 60601-1-2 Test
Level
Compatible Level
Electromagnetic Environment/Guidance
Electrostatic
discharge (ESD):
IEC 61000-4-2
±2, 4, 6 kV, contact
±2, 4, 8 kV, air
±2, 4, 6 kV, contact
±2, 4, 8 kV, air
Floors should be made of wood,
concrete or ceramic tiles. In a room
with a floor covered with synthetic
material, the relative humidity
should be at least 30%.
Electrically fast
transient/burst:
IEC 61000-4-4
±2 kV supply line
±1 kV I/O line
±2 kV supply line
±1 kV I/O line
The power supply should be of the
same quality as a commercial or
hospital line.
Surge:
IEC 61000-4-5
±0.5, 1 kV,
differential mode
±0.5, 1, 2 kV,
common mode
±0.5, 1 kV,
differential mode
±0.5, 1, 2 kV,
common mode
The power supply should be of the
same quality as a commercial or
hospital line.
Voltage drops,
dips or fluctuation
of input supply
line:
IEC 61000-4-11
<5% U
T
for 0.5 cycles
(>95% U
T
drop)
<40% U
T
for 5 cycles
(60% U
T
drop)
70% U
T
for 25 cycles
(>30% U
T
drop)
<5% U
T
for 5 seconds
(95% U
T
drop)
<5% U
T
for 0.5 cycles
(>95% U
T
drop)
<40% U
T
for 5 cycles
(60% U
T
drop)
70% U
T
for 25 cycles
(>30% U
T
drop)
<5% U
T
for 5 seconds
(95% U
T
drop)
The power supply should be of the
same quality as a commercial or
hospital line.
If continuous operation is required at
all times regardless of power
outages, connect the VS-1500 to an
uninterruptible power supply (UPS)
device.
Power frequency
magnetic fields
(50/60 Hz):
IEC 61000-4-8
3 A/m
3 A/m
The power frequency magnetic field
of the power supply should be the
same level as a standard
commercial or hospital line.
z
Note: U
T
is the AC line voltage prior to application of the test level.
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...