Chapter 5
Maintenance
Confidential
Copying Prohibited
5-43
●
Mechanical Inspection (
Perform the following inspection by actually operating the instrument.)
Item
Inspection Item
Procedure
Criteria
1
Keys
Verify that key operation is
smooth.
Operation should be smooth.
Main Unit
2
Recorder
Verify that recorder operation is
smooth, and no abnormal sound
is heard.
Operation should be smooth and
no abnormal sound should be
heard.
1
Power cable
Verify that it can be connected /
disconnected to the main unit
smoothly.
Smooth connection /
disconnection to the main unit
should be possible.
2 Lead
cable
Verify that it can be connected /
disconnected to the main unit
smoothly.
Smooth connection /
disconnection to the main unit
should be possible.
3 Clip
electrode
Verify proper connection to the
lead cable.
Should be properly connected.
4 Chest
electrode
Verify proper connection to the
lead cable.
Should be properly connected.
5 Cuff
Verify that it can be connected /
disconnected to the air hose
smoothly.
Smooth connection /
disconnection to the air hose
should be possible.
6 Air
hose
Verify that it can be connected /
disconnected to the main unit and
cuff smoothly.
Smooth connection /
disconnection to the main unit
and cuff should be possible.
7 PCG
microphone
Verify that it can be connected /
disconnected to the main unit
smoothly.
Smooth connection /
disconnection to the main unit
should be possible.
8
Amorphous PW
sensor
Verify that it can be connected /
disconnected to the main unit
smoothly.
Smooth connection /
disconnection to the main unit
should be possible.
Accessories
9
Air bag
pulse wave sensor
Verify that it can be connected /
disconnected to the main unit
smoothly, and sufficient air is
inside the air bag.
Smooth connection /
disconnection to the main unit
should be possible, sufficient air
should be inside the air bag.
●
Electrical Inspection
Item Inspection
Item
Procedure
Criteria
1 Power
supply
Turn the power ON and check the
display.
Display turns ON at power ON.
2 Display
Verify if any abnormality or flickering.
No abnormality or flickering
should be found. Waveforms
should be properly displayed.
3 Printout
Verify that waveforms can be printed.
Waveforms should be properly
printed.
4 Zero
balance
With each air connector open to the
atmosphere, check the displayed
pressure value.
Should be within
±
1mmHg.
Performanc
e
5 BP
measurement
Perform test measurement with cuffs
applied to both brachial and ankles.
Proper BP value should be
measured.
●
Maintenance Test
Inspection Item
Procedure
Criteria
Built-in recorder
test
Check according to “5.3.2 Testing
Built-in Thermal Array Recorder”.
There should be no zigzag, missing dots, uneven
prints. Recording speed should be within 25,
50mm ±3%.
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...